Abstract
Purpose :
OCT images often suffer from speckle noises and motion-induced artifacts that affect its accurate, reproducible, and reliable assessment of the samples To mitigate this challenge, we propose a method based on a fast, robust and three-dimensional registration method to enhance the contrast of both OCT and OCT-Angiography(OCT-A) images without pre-processing.
Methods :
All data were collected by a commercial SSOCTA device (PLEX® Elite 9000 (ZEISS, Dublin, CA)). 6x6 mm2 scans were utilized to test our method with a dimension 500x500x1560 (x, y, z). 10 repeated volumes were acquired on all participants. The 3D affine transformation was first implemented to low resolution images to correct the bulk motion between different OCT structural volumes and then use the non-ridge B-spline registration to refine the local alignment between the reference and moving volumes on both OCT structural and flow images. Finally, the corrected images were weight-averaged with the previous results and are assigned as the reference images for the next registration. The registration program was developed based on Matlab platform and it costs about 30 minutes for 10 volumes averaging on workstation configured with the Intel Xeon 2630-v3 CPU and 128G RAM.
Results :
Five subjects have been enrolled in this study with 10 repeated volume scans either with 3x3 mm2 scan or 6x6 mm2 scan. The deformation of sequential images can be rapidly corrected and the motion artifacts can be effectively suppressed. The average peak signal-to-noise ratio (PSNR) measures the misalignment of images, which increased from 25 dB to 72dB.
Conclusions :
The averaged results from both B-scan and the en-face projection of retinal layer and choriocapillaris showed an enhanced SNR and vascular connectivity comparing to the single scan image. This approach provides better visualization of microstructures and microvasculature that could be helpful in monitoring the disease progression and further assistant the therapeutic intervention.
This abstract was presented at the 2019 ARVO Imaging in the Eye Conference, held in Vancouver, Canada, April 26-27, 2019.