Comparison of dual 15x15 optical coherence tomography montage with fluorescein angiography fundus image on diabetic retinopathy patients
Author Affiliations & Notes
  • Sophie Kubach
    Carl Zeiss Meditec Inc, Dublin, California, United States
  • Jean-François KOROBELNIK
    Centre Hospitalier Universitaire de Bordeaux, Bordeaux, Aquitaine, France
  • Luis de Sisternes
    Carl Zeiss Meditec Inc, Dublin, California, United States
  • Footnotes
    Commercial Relationships   Sophie Kubach, Carl Zeiss Meditec, Inc (E); Jean-François KOROBELNIK, None; Luis de Sisternes, Carl Zeiss Meditec, Inc (E)
  • Footnotes
    Support  None
Investigative Ophthalmology & Visual Science August 2021, Vol.62, 25. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Sophie Kubach, Jean-François KOROBELNIK, Luis de Sisternes; Comparison of dual 15x15 optical coherence tomography montage with fluorescein angiography fundus image on diabetic retinopathy patients. Invest. Ophthalmol. Vis. Sci. 2021;62(11):25.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose : While conventional color fundus photography covers a wide area of the retina with a typical 90 deg field of view (FOV), fluorescein angiography (FA) is an invasive test that requires intravenous administration of dye. In comparison, optical coherence tomography angiography (OCTA) is a non-invasive imaging modality, however the FOV covered by OCTA imaging has remained limited in comparison with conventional color fundus photography. In this study, we compare clinical features present in the angiogram of diabetic retinopathy (DR) eyes obtained from 3D registration of a two 15x15 mm OCTA scans acquired at nasal and temporal fixation with FA images.

Methods : Two diabetic eyes were imaged on a CLARUS® 700 instrument using FA and on a PLEX® Elite 9000 (ZEISS, Dublin, CA) prototype instrument using the dual 15x15 volumetric montage protocol. FA images were acquired on CLARUS, while 2 individual 15x15 mm angio scans (about 55x55 deg) centered at nasal and temporal fixations were acquired on PLEX® Elite. The individual 15x15 OCTA volumes were registered to produce a 3D montaged scan covering a FOV of about 85 deg (temporal to nasal) by 50 deg (superior to inferior). Clinical features present in the angiograms were compared between the 2 imaging modalities.

Results : Both imaging modalities revealed new vascularization (Figure1) on an eye with moderate proliferative diabetic retinopathy (PDR). Leakage from pre-retinal vessels was observed on the FA image but their localizations above the ILM could only be revealed from the OCTA B-scans. Microaneurysms were better appreciated on the FA image, however regions of capillary dropout were clearly identified on the OCTA retina slab image and unresolved on the FA image. On another eye (Figure2) with severe PDR, vascular abnormalities were seen on both imaging modalities but with a higher contrast on the OCTA retinal slab image. Region of capillary dropout were clearly identified on the OCTA retinal slab image while remaining undetected on the FA image.

Conclusions : FA remains the best imaging modality to visualize microaneurysms and leakage, however OCTA is able to resolve low perfused areas that could otherwise remain undetected with FA. With OCTA scans approaching the FOV of fundus photography, the strength of these 2 imaging modalities are complementary in the early detection of retinal vascular abnormalities.