Our study may have some limitations that need to be considered. First, we chose to define glaucoma progression using a standard automated perimetry and a trend-based analysis with a linear regression of the mean deviation of visual field tests.
34 Retinal sensitivity values measured with standard automated perimetry could fluctuate for the same eye with increasing test-retest variability with decreasing retinal sensitivity values across the retina.
35 Furthermore, glaucoma progression can also be assessed using an event-based analysis of the visual field.
34 However, the trend-based analysis approach is commonly used in published studies analyzing glaucoma progression and also provides a rate of progression useful to analyze characteristics of glaucoma on a long-term period.
2,8,36 Furthermore, eyes were eligible when at least 5 reliable visual fields were performed over at least a period of 2 years and the mean number of visual field was 7.7 ± 2.3 in our population sample. At the beginning of standard of care follow-up visits, patients were classified as suspect of glaucoma progression on the visual field and progression was later confirmed with time and additional visual fields. Thus, we hypothesize that the rate of progression of our population sample was reliable and that our findings could be applied to such population groups. Another potential limitation could be related to the device we used to evaluate IOP related fluctuations and the potential bias that could be associated with this technology. Indeed, the 24-hour monitoring provided by the CLS measures the changes in corneal curvature and circumference expressed in voltage and thus does not provide real IOP measurements. Voltage measured with the strain gauge is supposed to be modified by changes in corneal circumference at the corneoscleral junction and the correlation between volumetric changes and IOP is not fully established. Hence, the correlation between voltage and IOP related fluctuations still remains unclear.
37,38 Whereas Mansouri et al. found that the coefficient of correlation between CLS and pneumatonometer was R
2 = 0.914, Vitish-Sharma et al. found that the mean correlation coefficient between CLS output signal measurements and IOP measurements was r = 0.291.
39,40 Additionally, the association of corneal parameters as corneal thickness at the corneal junction and CLS measurements or the influence of a 24-hour wearing of the CLS on the cornea could also influence the recorded CLS output signal particularly at the end of the monitoring.
29,31 In our population sample, we observed a 20-µm difference in mean CCT measurements between the two groups at baseline. Although the CLS measures the changes in corneal curvature or circumference at the corneal junction and not at the apex of the cornea, the influence of this difference in central corneal thickness on our results remains unclear. However, the score we calculated took into account central corneal thickness parameter and was still able to discriminate the two groups of progression. Finally, although our multivariate classifier model showed good diagnostic performances to diagnose OAG with a faster rate of progression, this classifier would also need to be tested in an independent and larger population sample to confirm or refine its diagnostic performances (external validity analysis) and enable generalization of findings.