There were 103 patients with anisometropic amblyopia (55 boys and 48 girls), who visited our hospitals between 1995 and 2021. Patients who had a hyperopic refractive error difference between the eyes of >1.50 diopters (D) or an astigmatic refractive error difference of >1.50 D with successfully treated visual acuity without manifest strabismus were included. Patients with accommodative esotropia, ocular disorders other than anisometropic amblyopia, neurologic disorders, cognitive impairment, or systemic diseases were excluded. Amblyopia was defined as the difference in the best-corrected visual acuity (BCVA) of worse than 0.0 log MAR in the affected eye and 0.2 log MAR difference between eyes. The detailed characteristics of all cases are mentioned in
Tables 1a and
b.
All patients were prescribed fully corrected glasses based on cycloplegic refraction. They were ordered to occlude the sound eye 2 hours daily if no fusional response occurred with Bagolini striated glasses. All patients underwent refractive adaptation before any occlusion. Amblyopia was resolved by achieving a BCVA of 0.0 log MAR or better in the previously amblyopic eye and less than 0.2 log MAR difference between eyes, and stabilization of this BCVA for 6 months or longer was considered as successful treatment.
Suppression scotomas were measured using P4D (see
Fig. 1a), a device which is a modification of the W4D. In P4D, 2 different polarization filters are attached to the device and glasses to perform binocular separation instead of the color filters used in the W4D device. This is because, in W4D, binocular color rivalry may be caused by looking through green and red filters with the left and right eyes, respectively.
There were 4 dots in front of the P4D device. The diameter of each dot was 4 cm. The distance from the middle of a dot to the center of the device was 8 cm. Dot A in
Figure 1a is nonpolarized. Dots B and C were polarized in the vertical direction, and dot D was polarized in the horizontal direction, using Wratten-type linear polarizing filters.
The patient wore polarizing filters in the form of glasses over their refractive correction lenses (see
Fig. 1b), which was corresponded to the same polarizing filters as the filters to the P4D device. The polarizing filter in the horizontal direction was corresponded to the glasses on the side of the amblyopic eye, and that on the side of the fellow eye was corresponded to the polarizing filter in the vertical direction (
Fig. 1c). Wearing glasses, the patient could see three dots (A, B, and C) with the fellow eye and two dots (A and D) with the amblyopic eye (see
Fig. 1c). Therefore, during a definite fusion response in both eyes of the patient, an individual could see four dots (A, B, C, and D). However, when suppression occurs in the amblyopic eye of the patient, one could only see three dots (A, B, and C).
The P4D device was set at 30 cm from the patient. The visual angle in prism diopters (PD) is 8 cm/0.3 m (i.e. 26.7 PD). The visual angle between the two opposite dots changes with changes in the distance between the patient and the device. The patient was instructed to wear the glasses, look at four dots on the device, and tell the examiner when the number of dots changed from four to three. The P4D device was moved slowly away from the patient until the four dots (i.e. a definite fusion response) changed to three dots (i.e. suppression in the amblyopic eye). The visual angle at the point where the dots changed from four to three corresponded to the suppression scotoma radius (i.e. the radius of the suppression size in PD, which is 8 cm/change distance in meters;
Fig. 2a). As shown in
Figure 2a, we can measure the lower radius of the suppression scotoma. The dots can be rotated by 360 degrees every 90 degrees and can be set to four position patterns (patterns ①, ②, ③, and ④ in
Fig. 2b). This enables the dot that the patient cannot see when the amblyopic eye is under suppression (i.e. see dot D in
Fig. 1a) to be placed in four different locations (see patterns ①, ②, ③, and ④ in
Fig. 2b). We can measure the upper radius in pattern ①, the right radius in pattern ②, the lower radius in pattern ③, and the left radius in pattern ④. We added right and left radii (i.e. the horizontal diameter of the suppression scotoma) and used them to calculate the effective suppression scotoma size in the anisometropic eye. The clinician measuring the size of the suppression scotoma was blind to the 4ΔBOT result. This quantitative measurement of the suppression scotoma was repeated three consecutive times in one patient, and the average was used for statistical analysis. To evaluate the reproducibility of this method of quantitative measurement of the suppression scotoma using P4D, we compared each of the 3 horizontal diameters of the suppression scotomas in 30 consecutive patients tested most recently from 103 patients included in this study. The intraclass correlation coefficient of this method was also evaluated.
Stereoacuity was evaluated using a Randot Stereotest (Stereo Optical Co., Chicago, IL, USA). As a mathematical rule, all stereoacuity values were converted to logarithmic values with a base of 20″, which is the minimum value of the Randot Stereotest, such as 20″ (1), 30″ (1.14), 40″ (1.23), 50″ (1.31), 70″ (1.42), 100″ (1.54), 140″ (1.65), 200″ (1.77), 400″ (2), and 500″ (2.07). Patients with no detectable stereoacuity (i.e. nil) were assigned a value of 7000″ (2.96). We investigated the correlations between logarithmic stereoacuity and horizontal diameter of suppression scotomas in all patients.
All patients were examined using the 4ΔBOT, a partial quantitative test used to detect the existence of a suppression scotoma larger than 4Δ. In this test, the patient was instructed to watch a fixation target. The examiner placed a 4Δ base-out prism in front of the patient's eye and observed the responses of both eyes. For a person who did not have a >4Δ suppression scotoma, the image was shifted by the prism and the eye had an inward movement.
Based on Herring's law of equal innervation, the fellow eye moves outward (phase 1). The patient simultaneously experienced diplopia and the fellow eye had a fusional convergence movement. This eye stopped at the convergence position (phase 2). The test was used to examine both the eyes of all patients several times. If the eye had a suppression scotoma of >4Δ when the examiner placed a 4Δ base-out prism in front of the patient's eye, the eye did not have an inward movement, and there were no other ocular responses (i.e. “no movement”). If the fellow eye had suppression scotoma, it had no phase 2 response (i.e. “no reverse”). Patients who demonstrated in phases 1 and 2 were defined as “normal responders.” Patients who often had a normal response but occasionally had no reverse response were defined as “subnormal responders.” All other responses were defined as “abnormal responders.” The clinician conducting the 4ΔBOT was blind to the stereopsis findings and size of the suppression scotoma in all cases. Based on these results, all patients were divided into 3 groups: normal responders, subnormal responders, and abnormal responders, comprising 29, 48, and 26 patients, respectively.
Statistical analyses were performed using the IBM SPSS Statistics for Windows, version 29.0 (IBM Corp., Armonk, NY, USA) and Microsoft Excel for Windows, version 16 (Microsoft Corp., Redmond, WA, USA). The horizontal diameter of the suppression scotoma, stereoacuity, pretreatment log MAR visual acuity, degree of anisometropia, and patient's age at the first visit to our hospital were compared among the three groups using 1-way ANOVA. The Bonferroni correction was applied to compare the three groups. The relationships between the size of the suppression scotomas and stereoacuity were investigated using Pearson's correlation coefficient. The significance level was adjusted to a family wise error of P < 0.017 for multiple comparisons. The tests were considered statistically significant when P < 0.05. We also evaluated the estimation accuracy for each linear relationship between the logarithmic stereoacuity and suppression scotoma size of the three groups using the root mean square error (RMSE) based on the statistical proof by SATISTA, Co., Ltd. (Uji, Kyoto, Japan).
This study was approved by the ethics committee of the Japan Community Health Care Organization Chukyo Hospital (Nagoya, Japan). Informed consent was obtained from all parents or guardians of the patients. All medical procedures were performed in accordance with the tenets of the Declaration of Helsinki.