A number of parameters in the lens are held constant in each minimization procedure and can have an influence on the final outcome. The influence of these parameters was evaluated by testing the minimization procedure on lens 1 for a range of values of these parameters and recording the minimum flens value and the corresponding average modulus Gmean.
Figure 6A shows the influence of the force boundary condition. For a range of force values from 10 to 80 mN, function
flens exponentially decreases in value from 42 µm for a force of 10 mN to a plateau of 12 µm for a force of 80 mN. The value of
Gmean obtained from the minimization procedure also increases with increasing force load. The interpretation of this result is that lower forces are not sufficient to deform the lens when accounting for capsular and zonular stiffness, even as
Gn and
Gc move toward zero. A similar effect is observed for the capsular modulus, represented in
Figure 6B. The average shear modulus of the lens decreases with increasing capsular modulus, along with the value of
flens.
Figure 6C shows the relationship between the capsule thickness and the estimated modulus. Per the slope of the graph, each additional micrometer of thickness decreases the average modulus by 0.225 kPa. The value of
flens was also found to decrease with decreasing capsular thickness. This could indicate that adjustments to the capsular thickness could improve the quality of the fit and minimization process. Although, in this case, the reduction in thickness improved the fit primarily in a region where the capsule is expected to be thickest.
Figure 6D shows the effect of altering the central thickness of the nucleus. The overall effects on the average modulus were low, but larger effects were seen in the nuclear modulus. The change from 2.2 to 2.6 mm of thickness resulted in an increase of 0.45 kPa in the estimated modulus of the nucleus.
Figure 6E shows the effect of altering the zonule spring constants. Data are presented for equatorial zonule constant (with anterior and posterior zonule constants both being multiples of this number). The nuclear/cortical moduli increased with increasing spring constant and
flens decreased. These effects indicate that the fixed parameters of the model affect the values of the estimated material properties and also affect the quality of fit, limiting their range of likely values.
In the model, forces of 60 mN were assumed.
Figure 6A shows that, beyond 50 mN, the value of
flens plateaus, indicating practically no further improvement in the quality of the fit. For forces below 50 mN, on the other hand,
flens is large and fit quality decreases. This suggests 50 mN is the lower boundary of the force during accommodation for these measurements. The upper boundary cannot be estimated from the change in
flens, but physiologically, the value is estimated to be around 80 mN.
17 Since force and accommodation are directly related, this suggests that stimuli of at least 4 D are required to properly estimate lens mechanical properties from patient data, at least in this age range.
In the model, several parameters were assumed to have an age dependence (
equations (1)–(
3)). As a relationship was found between the age of the participant and the moduli obtained from the minimization process, an additional test was done to assess the influence of the age-dependent parameters. A fixed value for each of these parameter (capsular modulus, ciliary body radius, zonule position) based on the average age of the participants (25.77 years) was used, and the minimization procedure was repeated for all lenses. This did not result in any substantial difference in correlation between age and average moduli (
R2 = 0.8) and resulted in an overall difference in average modulus of 0.23 kPa.