There was no statistically significant difference between the PCD or ICD of the baseline or last visit for each patient. Baseline mean (± SD) PCD was 16,079 ± 10,530 cones/mm
2, with a final visit PCD mean (± SD) of 15,974 ± 10,375 cones/mm
2; paired
t-test;
t(13) = 0.55,
P = 0.3. Baseline mean (± SD) ICD was 14.4 ± 5.9 µm, with a final visit ICD mean of 14.6 ± 6.15 µm, paired
t-test;
t(13) = –1.13,
P = 0.14. The estimated marginal mean PCD was calculated to be 15,346 cones/mm² at the mean age of 29 years old (SE = 2605), with a rate of change of –118 cones/mm² per year (SE = 76, conditional ICC = 0.98, df = 32, F = 2.4,
P = 0.13). The estimated marginal mean ICD was 13.8 µm (SE = 21.6), with a rate of change of 0.17 µm (SE = 0.75, conditional ICC = 0.99, df = 32, F = 0.49,
P = 0.83).
Figure 1 shows the PCD and ICD of each eye over the follow-up period, younger patients tended to have a greater change in PCD compared with older patients. Patient MM_0458 showed the largest absolute decrease in PCD of 9.2% over 4 years (22,441 to 20,601 cones/mm²).
Examination of the serial AOSLO imaging depicted in
Figure 2 qualitatively shows no obvious change over time.
Figure 3 shows examples of stable cone counts in 3 patients over the follow-up period in a region of interest. The repeatability of PCD measurements was high between the 2 measurements taken at each imaging visit with an ICC of 0.96 (95% confidence interval 0.90–0.98).